Body Material:PPS
Cable connection:Digital
Cable length:10 m
Compliance:Hazardous location, Maritime, CE
Electrode type:Specialty - GoldFlow:3 m (10 ft.) per second, maximum
Housing material:PPS
Material (electrode):Gold
Measuring range:-1500 to +1500 mV ORP
Mounting:Convertible
Operating temperature range:–5 to 70 °C (23 to 158 °F) pHD and ORP
Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers.
Sensitivity:± 0.5 mV
Sensor cable:Integral 4 conductor cable with one shield and polyurethane jacket; rated to 105 °C (221 °F); 10 m (33 ft.) standard length
Sensor type :Digital
Temperature accuracy:± 0.5 °C (± 0.9 °F)
Temperature sensor:Sensor
Pressure/Temperature Limits Digital: 6.9 bar
Transmission distance:1000 m (3280 ft.), maximum when used with a termination box 100 m (328 ft.), maximum
Warranty:24 months
Wetted materials:PEEK or PPS, salt bridge of matching material with PVDF junction, glass process electrode, titanium ground electrode, and FKM/FPM O-ring seals (pH sensor with optional HF-resistant glass process electrode has 316 stainless steel ground electrode, and perfluoroelastomer wetted O-rings; consult factory for other available wetted O-ring materials)